Skip to main content

Table 1 Data from easy laryngoscopy scenario.

From: Comparison of the Glidescope® and Pentax AWS®laryngoscopes to the Macintosh laryngoscope for use by Advanced Paramedics in easy and simulated difficult intubation

Parameter Assessed

Macintosh

Glidescope®

AWS®

P value

Overall Success Rate (%)

25 (100)

25 (100)

25 (100)

P = 1.0

Duration (sec, 1st attempt)

8 (7, 13)

11 (7, 17)

7 (6, 12)

P = 0.1

No of Intubation Attempts (%)

    

   1

24 (96)

24 (96)

25 (100)

P = 0.602

   2

1 (4)

1 (4)

0

 

   3

0

0

0

 

No of Optimization Maneuvers

    

   0

23 (92)

25 (100)

25 (100)

P = 0.368

   1

2 (8)

0

0

 

   > 1

0

0

0

 

Dental Compression [Severity]

    

   0

1 (4)*

11 (44)*

18 (72)*

P < 0.001

   Mild [+]

13 (52)

10 (40)

5 (20)

 

   Severe [++]

11 (44)

4 (16)

2 (8)

 
  1. Notes: Data are reported as mean ± SD, median (interquartile range) or as number (percentage).
  2. * Significantly [P < 0.05] different compared to both other groups