Table 2 Comparison of corrections after each route stage (by device)

From: Testing mechanical chest compression devices of different design for their suitability for prehospital patient transport - a simulator-based study

Device Correction of pressure point required?
No Yes
animax mono Number of route stages observed 3 15
Percentage (%) 17 83
AutoPulse Number of route stages observed 10 8
Percentage (%) 56 44
corpuls cpr Number of route stages observed 9 9
Percentage (%) 50 50
LUCAS 2 Number of route stages observed 16 2
Percentage (%) 89 11
\