Skip to main content

Table 3 Comparison of the 30:2 compression-ventilation ratio after each route stage (by device)

From: Testing mechanical chest compression devices of different design for their suitability for prehospital patient transport - a simulator-based study

Device

“30:2” ratio

Not OK

OK

animax mono

Number of route stages observed

9

6

Percentage (%)

60.0

40.0

AutoPulse

Number of route stages observed*

1

13

Percentage (%)

7.1

92.9

corpuls cpr

Number of route stages observed*

2

13

Percentage (%)

13.3

86.7

LUCAS 2

Number of route stages observed

2

13

Percentage (%)

13.3

86.7