Skip to main content

Table 3 Comparison of the 30:2 compression-ventilation ratio after each route stage (by device)

From: Testing mechanical chest compression devices of different design for their suitability for prehospital patient transport - a simulator-based study

Device “30:2” ratio
Not OK OK
animax mono Number of route stages observed 9 6
Percentage (%) 60.0 40.0
AutoPulse Number of route stages observed* 1 13
Percentage (%) 7.1 92.9
corpuls cpr Number of route stages observed* 2 13
Percentage (%) 13.3 86.7
LUCAS 2 Number of route stages observed 2 13
Percentage (%) 13.3 86.7
\