Skip to main content

Table 1 Devices and their maximum displacements (cranio-caudal; lateral)

From: Testing mechanical chest compression devices of different design for their suitability for prehospital patient transport - a simulator-based study

Device

Max. displacement (cm)

cranio-caudal

lateral

animax mono

4.5

1.0

AutoPulse

2.3

2.2

corpuls cpr

1.7

1.0

LUCAS 2

0.5

0.4