Impact of COVID-19-adapted guidelines using different airway management strategies on resuscitation quality in out-of-hospital-cardiac-arrest – a randomised manikin study

Abstract

Background

Although airway management for paramedics has moved away from endotracheal intubation towards extraglottic airway devices in recent years, in the context of COVID-19, endotracheal intubation has seen a revival. Endotracheal intubation has been recommended again under the assumption that it provides better protection against aerosol liberation and infection risk for care providers than extraglottic airway devices accepting an increase in no-flow time and possibly worsen patient outcomes.

Methods

In this manikin study paramedics performed advanced cardiac life support with non-shockable (Non-VF) and shockable rhythms (VF) in four settings: ERC guidelines 2021 (control), COVID-19-guidelines using videolaryngoscopic intubation (COVID-19-intubation), laryngeal mask (COVID-19-Laryngeal-Mask) or a modified laryngeal mask modified with a shower cap (COVID-19-showercap) to reduce aerosol liberation simulated by a fog machine. Primary endpoint was no-flow-time, secondary endpoints included data on airway management as well as the participants‘ subjective assessment of aerosol release using a Likert-scale (0 = no release–10 = maximum release) were collected and statistically compared. Continuous Data was presented as mean ± standard deviation. Interval-scaled Data were presented as median and Q1 and Q3.

Results

A total of 120 resuscitation scenarios were completed. Compared to control (Non-VF:11 ± 3 s, VF:12 ± 3 s) application of COVID-19-adapted guidelines lead to prolonged no-flow times in all groups (COVID-19-Intubation: Non-VF:17 ± 11 s, VF:19 ± 5 s;p ≤ 0.001; COVID-19-laryngeal-mask: VF:15 ± 5 s,p ≤ 0.01; COVID-19-showercap: VF:15 ± 3 s,p ≤ 0.01). Compared to COVID-19-Intubation, the use of the laryngeal mask and its modification with a showercap both led to a reduction of no-flow-time(COVID-19-laryngeal-mask: Non-VF:p = 0.002;VF:p ≤ 0.001; COVID-19-Showercap: Non-VF:p ≤ 0.001;VF:p = 0.002) due to a reduced duration of intubation (COVID-19-Intubation: Non-VF:40 ± 19 s;VF:33 ± 17 s; both p ≤ 0.01 vs. control, COVID-19-Laryngeal-Mask (Non-VF:15 ± 7 s;VF:13 ± 5 s;p > 0.05) and COVID-19-Shower-cap (Non-VF:15 ± 5 s;VF:17 ± 5 s;p > 0.05). The participants rated aerosol liberation lowest in COVID-19-intubation (median:0;Q1:0,Q3:2;p < 0.001vs.COVID-19-laryngeal-mask and COVID-19-showercap) compared to COVID-19-shower-cap (median:3;Q1:1,Q3:3 p < 0.001vs.COVID-19-laryngeal-mask) or COVID-19-laryngeal-mask (median:9;Q1:6,Q3:8).

Conclusions

COVID-19-adapted guidelines using videolaryngoscopic intubation lead to a prolongation of no-flow time. The use of a modified laryngeal mask with a shower cap seems to be a suitable compromise combining minimal impact on no-flowtime and reduced aerosol exposure for the involved providers.

Peer Review reports

Background

Based on experience with SARS-CoV-1, the resuscitation guidelines were adapted to mitigate the risk of infection for the rescuers, while accepting an increase in no-flow time and possibly worse patient outcomes [1,2,3,4,5,6,7,8]. Recent studies regarding out-of-hospital cardiac arrest (OHCA) demonstrated worse outcomes during the COVID-19 pandemic [9,10,11,12,13,14,15,16]. Potential explanations were lower survival rates caused by COVID-19, excessive utilization of health care capacities, prolonged response times due to donning of PPE, and the adaptations of resuscitation protocols recommending prolonged pauses in chest compressions for advanced airway management [1, 9,10,11, 17, 18]. Over the past years, prehospital airway management for paramedics has moved away from endotracheal intubation (ETI) towards supraglottic airway devices (SAD), skill acquisition and skill retention for ETI have proven difficult and there is evidence that SAD are non-inferior to ETI in OHCA [1,19.20,21]. In the context of COVID-19, however, ETI has made a comeback. It has been recommended when the condition is met that its complete seal of the trachea provides better protection against aerosol liberation and lowers infection risk for healthcare providers when compared to SAD or Bag-Mask-Ventilation during cardiopulmonary resuscitation (CPR) [1,2,3]. While studies have shown that the use of protective tents in the context of ETI led to a reduction in aerosol release, the effects of a protective film while using a SAD have not yet been investigated [22, 23]. Ideally, such protective film would allow the mouth-nose area to be sealed, leaving the neck, chest and extremities free for chest compressions, defibrillation and catheterisation. A modified SAD with a simple shower cap could therefore be suitable to elicit appropriate coverage. Due to the elastic band, it can be individually adjusted to the face of the wearer. The primary aim of the present study was to examine the impact of a laryngeal mask modified with a shower cap and other airway management strategies on no-flow-time and aerosol release against the background of the COVID-19 pandemic.

Methods

Ethical approval for this study (Ethical Committee No 2021-414-f-S) was granted on July, 2th 2021 by the Ethical Committee of the University Hospital of Muenster, Muenster, Germany (Chairperson Prof W.E. Berdel). The study was performed at a training center for paramedics in Bielefeld, Germany in August 2021 (‘Studieninstitut für kommunale Verwaltung Westfalen-Lippe, Fachbereich Medizin und Rettungswesen’). This manuscript adheres to the applicable CONSORT guidelines. After providing written informed consent, 60 paramedics were randomly allocated into 30 fixed teams, simulating the crew of an ambulance, and asked to perform a basic-life-support (BLS) OHCA scenario according to the guidelines of the European Resuscitation Council (ERC) 2021 [24]. All paramedics were Advanced Life Support trained prior to the inclusion. Before the scenario, the participants were intensely trained using manakin simulation training in the ERC COVID-19-adapted guidelines. The training focused on modified procedures of basic- and advanced-cardiac-life-support (detection of cardiac arrest, covering the face with an oxygen mask before starting chest compressions, early use of a defibrillator) and three different methods of airway-management with the intention to minimize aerosol release and maximize provider protection [1,2,3]: Performing a videolaryngoscopic intubation (COVID-19-Intubation), placing a LMA (COVID-19-Laryngeal-Mask) and placing a modified LMA with a shower cap attached, which was placed over the patients face in order to catch aerosol leaking from there the airway during CPR (COVID-19-Shower-cap; See Fig. 1). Supplement 1 shows the modification of the algorithm used (see Supplement 1). Instructions for constructing the modified LMA are shown in Supplement 2.

Fig. 1
figure 1

The laryngeal mask modified with a shower cap

Following the training, the previously formed teams were randomised via balanced randomization into two groups to investigate the effects on each initial rhythm: The first group completed the three scenarios of ERC COVID-19-adapted guidelines with a shockable, the second group with a non-shockable rhythm using each of the previously trained airway devices [2, 3]. To control for learning effects during the scenarios, the order of application of the different airway devices was systematically varied according to a Latin square. Figure 2 shows the group allocation process performed in the study.

Fig. 2
figure 2

Flowchart of the group allocation process

The target for the participants was to achieve the highest possible resuscitation quality with regard to protection specificities due to the COVID-19 pandemic [1,2,3]. The time required for airway management (intervals to application of oxygen, first airway management attempt, successful intubation/placement of the LMA, duration of intubation/LMA placement attempts, number of intubation/LMA placement attempts, unsuccessful intubation/LMA placement attempts, false intubations/LMA placement and first ventilation) as well as indicators of resuscitation quality according to current ERC guidelines (no-flow-time = time without chest compressions, total number of compressions, depth of compressions, compression frequency, proportion of compressions with correct hand-position and sufficient compression depth, time to first rhythm analysis and defibrillation) were measured [2, 3, 24].

The Laerdal Resusci Anne QCPR manikin (Laerdal Medical GmbH, Lilienthalstr.5, 82,178 Puchheim, Germany, http://www.laerdal.com/) was used. The participants’ equipment consisted of a fully equipped emergency backpack (“jump bag”), a ventilator (Medumat Standard²; Weinmann Emergency Medical Technology GmbH + Co. KG, Frohbösestraße 12, D-22,525 Hamburg, Germany, https://www.weinmann-emergency.com) and a transportable monitor-defibrillator (corpuls3, GS Elektromedizinische Geräte G. Stemple GmbH, Hauswiesenstraße 26, D-86,916 Kaufering, https://corpuls.world/). For videolaryngoscopic intubation the i-view TM videolaryngoscope was used (Intersurgical GmbH, Siegburger Straße 39, D-53,757 Sankt Augustin, https://de.intersurgical.com/info/iview); for SAD the LMA Classic size four (Teleflex Medical GmbH, Welfenstrasse 19, D-70,736 Fellbach, Germany, https://teleflex.com/emea/de/index), inflated with 30ml of air to reach a cuff pressure of 60cmH2O, as measured by universal cuff pressure measuring device (VBM Medizintechnik GmbH, Einsteinstrasse 1, D-72,172 Sulza.N., www.vbm-medical.de) was used.

To visualize the simulated aerosol release, the breathing system of the Laerdal Resusci Anne QCPR manikin was modified using the ‚Look Tiny Cx fog machine‘ (Look Solutions GmbH & Co. KG, Bünteweg 33, DE-30,989 Gehrden, https://www.looksolutions.com/) (See Supplement 3).

Data were transmitted via WLAN to the SimPad PLUS (SimPad PLUS, Laerdal Medical 2016, www.laerdal.com) with SkillReporter (Session Viewer, Laerdal Medical 2016, www.laerdal.com), preinstalled on the tablet. The individual datasets were saved and evaluated using the debriefing software Session Viewer 6.2.6400 by Laerdal.

The duration of each scenario was limited to 5 min. During the COVID-19 scenarios, participants wore PPE as recommended in the COVID-19-adapted guidelines (FFP3 mask (equivalent to the US N95 standard), eye and face protection, long-sleeved gown, and gloves) [1, 2, 5]. Importantly, donning the equipment was not part of the scenarios to ensure that only the effects of the algorithm changes were investigated. Previous studies suggest a timeframe of one to five minutes for donning PPE [1, 2, 5, 19].

After completion of all scenarios, the participants were asked to anonymously provide information on their subjective assessment of the amount of aerosol release (0 = no release – 10 = maximum release) and the feeling of safety using the different airway devices (safe, rather safe, rather unsafe, unsafe), the assessment of the quality of the different resuscitation scenarios with regard to the shortest possible no-flow time (good quality, rather good quality, rather bad quality, bad quality) and optimal protection of the rescuer (good protection, rather good protection, rather bad protection, bad protection).

The primary outcome was no-flow time, defined as time without chest compressions. Secondary outcomes were the established quality indicators of resuscitation as mentioned above as well as the results of the participants survey on rescuer safety.

Analysis

STATA version 16 (StataCorp. 2019. Stata Statistical Software: Release 16. College Station, TX: StataCorp LLC.) was used for statistical analyses. The sample size was calculated with a standardized tolerance limit of one standard deviation and no difference, 20 participants were required for a power = 0.8 with a significance level of 0.05. Mean difference between outcomes for the scenarios within the COVID-19 groups and compared to control group are adjusted for the correlation between measures obtained by the same team by using mixed models. We performed a non-inferiority analysis in which one scenario is said to be non-inferior to reference scenario if the lower limit of 95% confidence interval of the mean differences is less than half a standard deviation of the outcome for the reference scenario. The statistical evaluation of the provider survey was performed with the Friedman- and Wilcoxon–test. The significance level was set at P ≤ 0.05.

Results

Thirty different teams of paramedics (female = 39%, male = 61%) completed a total of 120 cardiac arrest scenarios apportioned equally to the four study groups.

Table 1 shows the results of circulatory and respiratory interventions. Compared to the control group (non-VF:11 ± 3 s; VF:12 ± 3 s), no-flow time was prolonged in the COVID-19-intubation (non-VF:17 ± 11 s,p ≤ 0.01; VF:19 ± 5 s, p ≤ 0.001), COVID-19-laryngeal-mask (non-VF:11 ± 4 s; VF:15 ± 5 s) and COVID-19-laryngeal-mask and -shower-cap (non-VF:11 ± 4 s; VF:15 ± 3 s) (each VF:p ≤ 0.01) groups. The results of the parameters for chest compressions are shown in supplement 4. The results of the comparison of COVID-19 groups with ERC-2021 are presented in Table 2. Table 3 contains the results of the comparison within the COVID-19 groups. None of the groups were non-inferior when compared to the control group.

Table 1 Circulatory and respiratory interventions
Table 2 Results of the non-inferiority trial COVID-19-groups vs. ERC-2021
Table 3 Results of the non-inferiority trial within COVID-19-groups

Participants ranked aerosol release higher when using the LMA (mean:7.1 ± 2.0; median:9;Q1:6;Q3:8) compared to the LMA with shower cap (mean:2.4 ± 1.5; median:3, Q1:1,Q3:3) and intubation (1.1 ± 1.4, median:0,Q1:1,Q3 3) (each p < 0.001 vs. laryngeal mask). Using the LMA with shower cap was subjectively worse than ETI (p < 0.001). Participants felt safer (p < 0.001 each) using both the LMA with shower cap and ETI compared to the LMA, but not when comparing the LMA with shower cap with endotracheal intubation (p > 0.05). Within the COVID-19 groups, participants reported significantly improved resuscitation quality and shorter no-flow time for the COVID-19-Laryngeal-mask group when compared to the COVID-19-intubation group (p < 0.001), but more aerosol exposure than for COVID-19-intubation (p < 0.001). The results of the survey of the participants regarding the different algorithms are shown in Table 4. Supplements 58 show the aerosol release of the different airway devices.

Table 4 Results of the survey

Discussion

The present paper compares no-flow time and established quality indicators of resuscitation as well as aerosol release using COVID-19 resuscitation guidelines including three different airway management strategies in accordance with the ERC-2021 using a simulation model [1,2,3, 24]. Compared to ERC-2021, the COVID-19-adapted guidelines using ETI led to a prolongation of the no-flow time when compared to SAD. The use of modified laryngeal mask with a shower cap had minimal impact on no-flow time and markedly reduced aerosol exposure for the involved providers.

Over the last decade, the impact of different airway management strategies on outcomes following OHCA has been investigated in various studies [19, 20]. Although prolonged intubation attempts in particular were associated with a prolongation of the no-flow time, potentially leading to a delay in the rate of ROSC with poorer outcome [19, 20, 22, 24], ETI was explicitly recommended in the guidelines for COVID-19 resuscitation as the gold standard for reducing aerosol release in view of the idea of optimal protection of the rescuers [1, 2]. The present study supports previous data showing a prolongation of no-flow time during resuscitation even with the use of videolaryngoscopy and can thus at least partially explain the worsening of patients´ outcome at the time of the COVID-19 pandemic indicating that these need to be optimized in the light of future pandemics [1, 3, 9, 10, 16].

Since the insertion of SAD is easy to both learn and put into practice for non-medical emergency service personnel, their use is widespread in Europe. Due to the relevant risk of release of infectious aerosols, airway interventions in patients with COVID-19 are associated with a significant risk of infection. Especially during face mask and supraglottic airway ventilation, intubation, extubation, and cardiopulmonary resuscitation, an increased aerosolization could be observed [25,26,27,28]. Nevertheless, in COVID-19 resuscitation, the use of SAD by rescuers not experienced in ETI was recommended knowing well the higher risk of aerosol release [1, 3]. To minimise virus transmission, various studies investigated different barrier systems, such as intubation tents, bag barrier drape systems etc., showing that containment systems can reduce the spread of infectious respiratory secretions during simulated coughing or extubation [26,27,28]. Despite these advantages, these systems had disadvantages in resuscitation situations, such as the time required to set up the device for patient use, covering the airways as quickly and easily as possible, easy access to the airway device, containment of aerosolization, difficult access to the patient to place defibrillation electrodes or catheters, e.g. puncture of the internal jugular vein, or problems during patient transport, as not every intubation tent can be easily attached to an ambulance stretcher [26,27,28]. The present pilot study suggests that the use of a modified LMA with a shower cap may be a viable alternative for airway management in the context of a pandemic with droplet and/or aerosol transmissible pathogens [3]: Firstly, deteriorations regarding no-flow time and total number of chest compressions were only observed in the VF group, making the COVID-19-laryngeal-mask- and COVID-19-shower-cap-group the best performing section compared to the other study groups. Secondly, participants rated aerosol release as lower and rescuer safety as higher compared to the LMA without shower cap. Thirdly, the production of an appropriately modified airway is simple, inexpensive and would therefore be easy to implement even in low-income countries where vaccination strategies may not have been implemented across the board or in anti-vaccine countries. Fourthly, such a modified LMA would allow the mouth-nose area to be sealed, leaving the neck, chest and extremities free for chest compressions, defibrillation and catheterization. Finally, the relative ease of use of SAD has been curbed by concerns regarding an incomplete seal of the airways causing an increased risk of liberation of infectious aerosols. This is especially so in dynamic resuscitation scenarios with risk of dislocation of the airway device and when the cuff pressure was limited to 60 cmH20 as in the present study [3, 28,29,30]: The use of a modified LMA with a shower cap seems to be the optimal compromise between helper safety on the one hand and optimal resuscitation quality on the other [2, 30].

SAD manufacturers should consider incorporating the shower cap concept into their SAD and prospective studies should evaluate this concept with regard to complications such as obstructing access to the airway in situations such as vomiting, an additional risk be a unnoticed tear in the shower cap, which may lead to a false sense of safety, different head-shapes, patients with a large beard or additional wasted plastic for times when an SAD is placed but CPR is not being performed. In addition, an SAD modified with a shower cap could represent a further building block not only in the context of prehospital resuscitation or difficult airway protection but also for anaesthesia.

Limitations

As the influence of the different airway devices on key performance indicators of resuscitation was investigated, the application of the PPE was not part of the study, but obviously leads to a further prolongation of no-flow time [1, 17]. Performing resuscitation under observation during the study may have influenced the participants (Hawthorne- effect). Furthermore, manikin-based investigations provide only a limited reproduction of reality e.g. the patients’ anatomy when inserting a SAD and patients` outcomes cannot be evaluated. The present study investigated aerosol liberation when a LMA classic is employed. Different kinds of SAD such as the laryngeal tube may have different leak pressures. However, it can be assumed that the use of a shower cap will lead to a reduction in aerosol exposure with other SAD also. Further studies are necessary to evaluate the influence of a shower cap in combination with different SAD. In addition, in the present study, simulation of aerosol liberation was simulated by means of a fog machine. Aerosol dispersion may not fully reflect reality. Quantity of aerosol release could not be measured objectively. Evaluation on the basis of the participants assessment has limitations. Currently, there are other methods to measure the aerosol release objectively, such as digital analysis of video recordings or optical particle sizer. Nevertheless, with regard to the heterogenous OHCA patients, a simulation model may provide standardized and valuable insights on optimum airway management in OHCA in the light of the COVID-19 pandemic. As a result, the observed effects may be of significance in view of the current situation with the spread of the COVID-19 pandemic and the expected deterioration in the quality of resuscitation because there are only limited possibilities for the urgently required clinical evaluation of the corresponding guideline changes. In addition, results for the COVID-19-adapted guidelines may improve following more intensified training of the providers under adequate hygiene concepts. Possibly non-inferiority could not be proven due to an insufficient number of participants. However, no-flow time did not differ between the LMA and shower cap groups, whereas aerosol release was rated lower so that an increase in the number of cases could prove non-inferiority.

Conclusions

The present study shows that the COVID-19-adapted guidelines using ETI led to a prolongation of the no-flow time, which markedly worsened the overall quality of resuscitation. These effects can be attenuated using an SAD. Although the influence of airway management on the outcome of resuscitation and the transmission of Sars-CoV-2 is critically discussed, the use of a modified LMA with a shower cap seems to be suitable to ensure optimal resuscitation quality and to reduced aerosol exposure at the same time.

Data Availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

BLS:

Basic Life-Support

CPR:

Cardiopulmonary Resuscitation

COVID-19:

Corona-Virus-Disease 2019

ETI:

Endotracheal Intubation

ERC:

European Resuscitation Council 2021

EGA:

Extraglottic Airway Devices

FFP3:

Filtering Face Piece 3

LMA:

Laryngeal Mask Airway

OHCA:

Out-of-Hospital Cardiac Arrest

PPE:

Personal Protective Equipment

SARS-CoV-1:

Severe Acute Respiratory Syndrome Corona-Virus 1

US:

United States

VF:

Ventricular Fibrillation

References

  1. Jansen G, Ebeling N, Latka E, Krüger S, Scholz SS, Trapp S, et al. Impact of COVID-19-adapted guidelines on resuscitation quality in out-of-hospital-cardiac-arrest: a manikin study. Minerva Anestesiol. 2021;87:1320–9.

    Article  PubMed  Google Scholar 

  2. Perkins GD, Morley PT, Nolan JP, Soar J, Berg K, Olasveengen T, et al. International Liaison Committee on Resuscitation: COVID-19 consensus on science, treatment recommendations and task force insights. Resuscitation. 2020;151:145–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Nolan JP, Monsieurs KG, Bossaert L, Böttiger BW, Greif R, Lott C, et al. European Resuscitation Council COVIDGuideline writing groups. European Resuscitation Council COVID-19 guidelines executive summary. Resuscitation. 2020;153:45–55.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Couper K, Taylor-Phillips S, Grove A, Freeman K, Osokogu O, Court R, et al. COVID-19 in cardiac arrest and infection risk to rescuers: a systematic review. Resuscitation. 2020;151:59–66.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Raboud J, Shigayeva A, McGeer A, Bontovics E, Chapman M, Gravel D, et al. Risk factors for SARS transmission from patients requiring intubation: a multicentre investigation in Toronto, Canada. PLoS ONE. 2010;5:e10717.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Brown E, Chan LM. Should chest compressions be considered an aerosol-generating procedure? A literature review in response to recent guidelines on personal protective equipment for patients with suspected COVID-19. Clin Med (Lond). 2020;20:e154–9.

    Article  PubMed  Google Scholar 

  7. Malysz M, Dabrowski M, Böttiger BW, Smereka J, Kulak K, Szarpak A, et al. Resuscitation of the patient with suspected/confirmed COVID-19 when wearing personal protective equipment: a randomized multicenter crossover simulation trial. Cardiol J. 2020;27:497–506.

    PubMed  PubMed Central  Google Scholar 

  8. Hwang SY, Yoon H, Yoon A, Kim T, Lee G, Jung KY, et al. N95 filtering facepiece respirators do not reliably afford respiratory protection during chest compression: a simulation study. Am J Emerg Med. 2020;38:12–7.

    Article  PubMed  Google Scholar 

  9. Lim ZJ, Ponnapa Reddy M, Afroz A, Billah B, Shekar K, Subramaniam A. Incidence and outcome of out-of-hospital cardiac arrests in the COVID-19 era: a systematic review and meta-analysis. Resuscitation. 2020;157:248–58.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Scquizzato T, Landoni G, Paoli A, Lembo R, Fominskiy E, Kuzovlev A, et al. Effects of COVID-19 pandemic on out-of-hospital cardiac arrests: a systematic review. Resuscitation. 2020;157:241–7.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Shao F, Sun P, Tang Z. Cardiopulmonary resuscitation of inpatients with severe COVID-19 pneumonia: the Wuhan experience. Resuscitation. 2020;152:95–6.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Thapa SB, Kakar TS, Mayer C, Khanal D. Clinical outcomes of inhospital cardiac arrest in COVID-19. JAMA Intern Med. 2020;181:e204796.

    Google Scholar 

  13. Chan PS, Girotra S, Tang Y, Al-Araji R, Nallamothu BK, McNally B. Outcomes for out-of-hospital cardiac arrest in the United States during the coronavirus disease 2019 pandemic.JAMA Cardio. 2021;6:296–303.

  14. Baert V, Jaeger D, Hubert H, Lascarrou JB, Debaty G, Chouihed T, et al. Assessment of changes in cardiopulmonary resuscitation practices and outcomes on 1005 victims of out-of-hospital cardiac arrest during the COVID-19 outbreak: registry-based study. Scand J Trauma Resusc Emerg Med. 2020;28:119.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Baldi E, Sechi GM, Mare C, Canevari F, Brancaglione A, Primi R, et al. Treatment of out-of-hospital cardiac arrest in the COVID-19 era: a 100 days experience from the Lombardy region. PLoS ONE. 2020;15:e0241028.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Ball J, Nehme Z, Bernard S, Stub D, Stephenson M, Smith K. Collateral damage: hidden impact of the COVID-19 pandemic on the out-of-hospital cardiac arrest system-of-care.Resuscitation. 2020;156:157–63.

  17. Watson L, Sault W, Gwyn R, Verbeek PR. The “delay effect” of donning a gown during cardiopulmonary resuscitation in a simulation model. CJEM. 2008;10:333–8.

    Article  PubMed  Google Scholar 

  18. Chen J, Lu KZ, Yi B, Chen Y. Chest compression with personal protective equipment during cardiopulmonary resuscitation: a randomized crossover simulation study. Med (Baltim). 2016;95:e3262.

    Article  Google Scholar 

  19. Benger JR, Kirby K, Black S, Brett SJ, Clout M, Lazaroo MJ, et al. Effect of a strategy of a Supraglottic Airway device vs tracheal intubation during out-of-hospital Cardiac arrest on functional outcome: the AIRWAYS-2 Randomized Clinical Trial. JAMA. 2018;320:779.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Wang HE, Schmicker RH, Daya MR, Stephens SW, Idris AH, Carlson JN, et al. Effect of a strategy of initial laryngeal tube insertion vs endotracheal intubation on 72-Hour survival in adults with out-of-hospital cardiac arrest: a Randomized Clinical Trial. JAMA. 2018;320:769.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Andersen LW, Granfeldt A, Callaway CW, Bradley SM, Soar J, Nolan JP, et al. Association between Tracheal Intubation during Adult In-Hospital cardiac arrest and survival. JAMA. 2017;317:494–506.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Yang YL, Huang CH, Luk HN, Tsai PB. Adaptation to the plastic barrier sheet to facilitate Intubation during the COVID-19 pandemic. Anesth Analg. 2020;131:e97–9.

    Article  CAS  PubMed  Google Scholar 

  23. Fang PH, Lin YY, Lin CH. A Protection Tent for Airway Management in patients with COVID-19 infection. Ann Emerg Med. 2020;75:787–8.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Olasveengen TM, Semeraro F, Ristagno G, Castren M, Handley A, Kuzovlev A, et al. European Resuscitation Council Guidelines 2021: Basic Life Support. Resuscitation. 2021;161:98–114.

    Article  PubMed  Google Scholar 

  25. Ott M, Milazzo A, Liebau S, Jaki C, Schilling T, Krohn A, et al. Exploration of strategies to reduce aerosol-spread during chest compressions: a simulation and cadaver model. Resuscitation. 2020;152:192–8.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Liu Y, Yan LM, Wan L, Xiang TX, Le A, Liu JM, et al. Viral dynamics in mild and severe cases of COVID-19. Lancet Infect Dis. 2020;20:656–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Canelli R, Connor CW, Gonzalez M, Nozari A, Ortega R. Barrier enclosure during endotracheal intubation. N Engl J Med. 2020;382:1957–8.

    Article  PubMed  Google Scholar 

  28. Matava CT, Yu J, Denning S. Clear plastic drapes may be effective at limiting aerosolization and droplet spray during extubation: implications for COVID-19. Can J Anaesth. 2020;67:902–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Cameron AE, Sievert J, Asbury AJ, Jackson R. Gas leakage and the laryngeal mask airway. A comparison with the tracheal tube and facemask during spontaneous ventilation using a circle breathing system. Anaesthesia. 1996;51:1117–9.

    Article  CAS  PubMed  Google Scholar 

  30. Hsu A, Sasson C, Kudenchuk PJ, Atkins DL, Aziz K, Becker LB, et al. 2021 Interim Guidance to Health Care Providers for Basic and Advanced Cardiac Life support in adults, children, and neonates with suspected or confirmed COVID-19. Circ Cardiovasc Qual Outcomes. 2021;14:e008396.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Marvin Deslandes for revision and Helge Myklebust and Laerdal Medical for kindly providing the ’Shower-cap’ concept.

Funding

No funding was obtained for this study.

Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

SSS has been involved in acquisition of data, drafting the manuscript, revising it critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. SL has been involved in acquisition of data, drafting the manuscript, revising it critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. EL has been involved in acquisition of data, revising the manuscript critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. TB has been involved in acquisition of data, revising the manuscript critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. DK has been involved in acquisition of data, revising the manuscript critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. DT has been involved in acquisition of data, revising the manuscript critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. MW has been involved in acquisition of data, revising the manuscript critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. OS has made substantial contributions to conception and design, analysis and interpretation of data, revising the manuscript critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. SWR has made substantial contributions to conception and design, interpretation of data, drafting the manuscript and revising it critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. KCT has made substantial contributions to conception and design, interpretation of data, drafting the manuscript and revising it critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. GJ has made substantial contributions to conception and design, acquisition and interpretation of data, drafting the manuscript and revising it critically for important intellectual content, gave final approval of the version to be published and gave the Agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Corresponding author

Correspondence to Gerrit Jansen.

Ethics declarations

Ethics approval and consent to participate

Ethical approval for this study (Ethical Committee No 2021-414-f-S) was granted on July, 2th 2021 by the Ethical Committee of the University Hospital of Muenster, Muenster, Germany (Chairperson Prof W.E. Berdel). All methods were carried out in accordance with relevant guidelines and regulations in Ethics declaration section. Informed consent was obtained from all participants.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Scholz, S.S., Linder, S., Latka, E. et al. Impact of COVID-19-adapted guidelines using different airway management strategies on resuscitation quality in out-of-hospital-cardiac-arrest – a randomised manikin study. BMC Emerg Med 23, 48 (2023). https://doi.org/10.1186/s12873-023-00820-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12873-023-00820-y

Keywords